NAME
       xoscope - Digital Oscilloscope

SYNOPSIS
       xoscope [X toolkit options] [xoscope options] [file]

DESCRIPTION
       Xoscope	is  a  digital real-time oscilloscope. It graphically displays
       signal amplitude	or bit logic as	a function of time.   Signals  may  be
       displayed,  saved, recalled, and	manipulated by math functions.	Signal
       input devices currently include:

       /dev/dsp
	    Audio sound	recording via /dev/dsp.	 Two 8-bit analog channels  at
	    8000 S/s to	44100 S/s.  Left and right audio is connected to A and
	    B  inputs  respectively.   Use an external mixer program to	select
	    which sound	inputs to record.  AC coupled, voltages	unknown,  256K
	    sample memory.

       EsounD
	    Shared  audio  sound  via  the  Enlightened	Sound Daemon.  This is
	    great for watching music but support for it	is an option  at  com-
	    pile-time.	EsounD is auto-detected	and preferred over /dev/dsp.

       ProbeScope / OsziFOX
	    Radio  Shack ProbeScope, Cat. No. 22-310 is	also known as an oszi-
	    FOX.  This handheld	probe sends its	data through  a	 serial	 port.
	    It samples one channel at 6-bits up	to 20 MS/s with	128 samples of
	    memory.  Real voltages are labeled in sample ranges	from 1 volt to
	    100	 volts.	 If a ProbeScope is detected, it is connected to the A
	    input.

       Bitscope
	    Bitscope (www.bitscope.com)	is a mixed-signal capture engine which
	    is accessed	through	a serial port.	It  simultaneously  samples  a
	    digital  8-bit port	and two	analog channels	at 8 bit resolution at
	    up to 25 MS/s or more.  If detected, Channel A and B are connected
	    to X and Y while the Logic Analyzer	is connected to	 C.   Bitscope
	    support  is	 currently  under  development and not yet fully func-
	    tional.

       COMEDI
	    The	COMEDI project (www.comedi.org)	develops Linux drivers,	tools,
	    and	libraries for data acquisition.	 Many  commercially  available
	    ADC	cards are supported by COMEDI, and Xoscope can receive signals
	    from them via the COMEDI library.

	    See	 the  -x  and -z options and the ENVIRONMENT section below for
	    more details on how	the above devices are detected.	 Some  of  the
	    controls  below  apply  only  to the sound card and	are labeled as
	    such.  Xoscope has no physical control over	the ProbeScope/osziFOX
	    which is controlled	 by  its  own  switches	 and  built-in	menus.
	    Please  refer to your ProbeScope or	osziFOX	Owner's	Manual for op-
	    erating instructions.   Bitscope  will  eventually	be  controlled
	    through a separate dialog window.

RUN-TIME KEYBOARD CONTROLS
       Xoscope is an interactive program and can be completely controlled from
       the keyboard at run-time.  In verbose key help mode, each available key
       is shown	on the screen in (parentheses).	 The following single key com-
       mands are available:

       ?    Toggle verbose key help display mode.

       Escape
	    Immediately	quit the program.

       @    Load a previously saved file.  You are prompted for	the filename.

       #    Save  current  settings  and  memory buffers to a file that	can be
	    loaded later.  You are prompted for	the  filename  and  asked  for
	    confirmation to overwrite if it already exists.

       Enter
	    Clear and refresh the entire screen.

       &    Cycle  between the various input devices.  Note that this key will
	    not	toggle to an unresponsive input	device,	so if only one	device
	    is present,	it will	appear to have no effect.

       *    Different behavior for different input devices

	    Under EsounD, this value instead determines	whether	the connection
	    to EsounD will block or not.  Blocking mode	is nicest to CPU usage
	    but	 the  xoscope  interface will not respond when the there is no
	    sound stream coming	from EsounD.  Nonblocking mode	will  let  xo-
	    scope  be  responsive  whether sound is available or not, but will
	    consume all	available CPU cycles.

	    Under COMEDI, this key toggles between different analog  reference
	    points (ground, differential, or common).

       ^    Different behavior for different input devices

       (/)  Decrease/increase the sampling rate.

       9/0  Increase/decrease  the  Sec/Div horizontal time scale (zoom	out/in
	    on time).

       -/=  Decrease/increase the trigger level.

       _    Cycle the trigger channel.

       +    Cycle the trigger type: none, rising edge, or falling edge.

       Space
	    Cycle the trigger mode: run, wait, stop.   Run  mode  continuously
	    acquires  and  displays  samples  after trigger events.  Wait mode
	    waits for the first	trigger	event and displays only	the first  set
	    of	samples;  this	is "single-shot" mode.	Stop mode suspends the
	    data acquisition and displays the current samples.

       !    Cycle the plotting mode: point, point accumulate,  line,  or  line
	    accumulate.	  In  the  accumulate  modes,  all samples stay	on the
	    screen; use	Enter to clear them.

       ,    Cycle the graticule	style: none, minor divisions  only,  or	 minor
	    and	major divisions.

       .    Toggle the graticule position: behind or in	front of the signals.

       '    Toggle  the	 manual	 cursors on/off.  When manual cursors are dis-
	    played, the	measurements between the cursor	positions  are	shown.
	    When cursors are not displayed, automatic measurements are shown.

       "    Reset  both	manual cursor positions	to the sample just after trig-
	    ger.

       Ctrl-q/w/e/r
	    The	Control	key held down in combination with  q/w/e/r  moves  the
	    first cursor back or forward by 10 samples or back or forward by 1
	    sample respectively.

       Ctrl-a/s/d/f
	    The	 Control  key  held down in combination	with a/s/d/f moves the
	    second cursor back or forward by 10	samples	or back	or forward  by
	    1 sample respectively.

       1-8  Select  the	 corresponding display channel.	 Measurements are dis-
	    played for the channel.  Channel 1 and 2 are used as input to  the
	    math functions so they can't be used to do math.  By default, they
	    are	 connected to the A and	B input	channels.  Channel 1 and 2 can
	    also be used to display memory buffers or for doing	math on	memory
	    or the alternate input.  Channel 3 through 8  are  not  restricted
	    and	 can  be  used for any purpose.	 The remaining single key com-
	    mands operate on the currently selected channel:

       Tab  Toggle visibility: Hide or show the	selected channel.

       {/}  Decrease/Increase vertical scale of	the selected channel.

       [/]  Decrease/Increase vertical position	of the selected	channel.

       `/~  Decrease/Increase number of	logic analyzer	bits  displayed.   The
	    default  of	zero bits plots	the signal as one analog line of vary-
	    ing	amplitude.  Any	other value plots multiple digital lines  rep-
	    resenting the least	significant bits from bottom to	top.

       ;/:  Increase/Decrease the math function	of the selected	channel.  This
	    is not available on	channel	1 & 2.

       $    Show  the result of	an external math command on the	selected chan-
	    nel.  You are prompted for the command.  The command  must	accept
	    samples  of	 channel 1 & 2 on stdin	and write a new	signal to std-
	    out.  See operl, offt.c and	xy.c in	the distribution for  examples
	    of external	math filter commands.  Not available on	channel	1 & 2.

       a-z  Recall the corresponding memory buffer or input device to the cur-
	    rently  selected channel.  Input device channels are mapped	to the
	    earliest letters of	the alphabet; the  rest	 of  the  buffers  are
	    available for signal memory.

       A-Z  Store the currently	selected channel into the corresponding	memory
	    buffer.   Early  letters  of  the alphabet can not be used because
	    they're reserved as	the signal inputs,  so	the  exact  number  of
	    available  buffers is dependant on the input device.  Memories are
	    stored from	time zero to the current display update	position.   So
	    it is best to STOP the display before storing to a memory buffer.

MOUSE CONTROLS
       Xoscope	adds  mouse controls to	menus or around	the edges of the scope
       area.  These should be nearly self-explanatory.	They perform the  same
       functions  as  the  equivalent  keyboard	commands above.	 If built with
       GTK+, a context-sensitive pop-up	menu is	available with right-click  to
       select  channels,  change  scale	and position, recall and store signals
       and so on.  Left	click decreases	 a  variable  while  right  click  in-
       creases.	  The  manual  measurement cursors can also be positioned with
       the mouse.

COMMAND-LINE OPTIONS
       The command-line	options	define the startup state of xoscope  and  have
       reasonable  defaults.  All options may be capitalized in	case they con-
       flict with an X toolkit option.	These options  are  also  recorded  in
       text files saved	by xoscope.

       -h   Help  usage	 message  showing these	startup	options	with their de-
	    fault values, then exit.

       -# <code>
	    Startup conditions of each channel.	 # is a	channel	number from  1
	    to	8.  Code can have up to	three fields, separated	by colons: po-
	    sition[.bits][:scale[:function #, memory letter, or	external  com-
	    mand]].   Position is the number of	pixels above (positive)	or be-
	    low	(negative) the center of the display.  Bits is the  number  of
	    logic  analyzer  bits to display.  Scale is	a valid	scaling	factor
	    from 1/50 to 50, expressed as a fraction.	The  third  field  may
	    contain  a built-in	math function number, memory letter, or	exter-
	    nal	math command to	run on the channel.  Using these options makes
	    the	channel	visible	unless position	begins with a  '+',  in	 which
	    case the channel is	hidden.

       -a <channel>
	    Active, or selected, channel.

       -r <rate>
	    Sampling  Rate in samples per second.  For the sound card, current
	    valid values are 8000, 11025, 22050, or 44100.

       -s <scale>
	    Time Scale factor from 1/20	to 1000	expressed as a fraction	 where
	    1/1	is 1 ms/div.

       -t <trigger>
	    Trigger  conditions.   Trigger  can	have up	to three fields, sepa-
	    rated by colons: position[:type[:channel]].	 Position is the  num-
	    ber	 of  pixels above (positive) or	below (negative) the center of
	    the	display.  Type is a number indicating the kind of trigger, 0 =
	    automatic, 1 = rising edge,	2 = falling edge.  Channel should be x
	    or y.

       -l <cursors>
	    Manual cursor Line	positions.   Cursors  can  have	 up  to	 three
	    fields,  separated	by colons: first[:second[:on?]].  First	is the
	    sample position of the first cursor.  Second is the	 sample	 posi-
	    tion  of the second	cursor.	 The final field is weather the	manual
	    cursors are	displayed (1) or the not displayed (0).

       -p <type>
	    Plot type.	0 = point, 1 = point accumulate, 2 = line,  3  =  line
	    accumulate,	4 = step, 5 = step accumulate.

       -g <style>
	    Graticule  style.	0  = none, 1 = minor divisions only, 2 = minor
	    and	major divisions.

       -b   Whether the	graticule is drawn Behind or in	front of the signals.

       -v   Whether the	Verbose	key help is displayed.

       -x   Whether the	sound card input device	(XY) is	turned on.   This  can
	    be used to skip the	attempt	to connect to Esound or	/dev/dsp.

       -z   Whether  the  serial  input	 device	(Z) is turned on.  This	can be
	    used to suppress the search	for a serial scope device.

       file The	name of	a file to load upon startup.  This should  be  a  file
	    previously saved by	xoscope.

EXAMPLES
       xoscope -1 80 -2	-80 -3 0:1/5:6 -4 -160:1/5:7

	    This  runs	xoscope	 with  channel 1 above and channel 2 below the
	    center of the display.  Also channel 3 and 4 are made  visible  to
	    show the FFT of channel 1 and 2 respectively at a reduced scale of
	    1/5.

       xoscope oscope.dat

	    This runs xoscope, loading settings	and memory buffers from	a pre-
	    viously saved data file called "oscope.dat".

FILES
       Xoscope creates readable	text data files.  The files contain at least a
       comment	header	which  holds the current settings of xoscope.  Loading
       the file	causes these saved settings to be restored.

       To record  your	signals	 permanently  first  store  them  into	memory
       buffers,	 optionally  recall  them to channels, and then	save the file.
       All non-empty memory buffers are	written	to a column of the  file  fol-
       lowing  the  comment  header.  Columns are separated by tab characters.
       These are stored	back into the memory buffers when the  file  is	 later
       loaded.	Simply recall them to channels to view them.

       This  format  could  also  be read by some spreadsheet or plotting pro-
       grams.  For example, the	gnuplot	(1) command

       plot "oscope.dat" using 0:1, "oscope.dat" using 0:2

       would plot the first and	second columns of the "oscope.dat" data	file.

ENVIRONMENT
       OSCOPEPATH
	    The	path to	use when looking for external math commands.   If  un-
	    set, the built-in default is used.

       PROBESCOPE
	    The	 serial	device your ProbeScope or osziFOX is connected to.  If
	    unset, /dev/probescope is used.  /dev/probescope could be  a  sym-
	    bolic link to the real device such as /dev/ttyS1.

       BITSCOPE
	    The	 serial	 device	 your  Bitscope	 is  connected	to.  If	unset,
	    /dev/bitscope is used.  /dev/bitscope could	be a symbolic link  to
	    the	real device such as /dev/ttyS1.

       ESPEAKER
	    The	 host:port  of	the  EsounD to connect to if built with	EsounD
	    support.  If unset,	localhost is assumed.  If no EsounD connection
	    is made or if there	is no EsounD support compiled in, then xoscope
	    will try to	read /dev/dsp directly.

LIMITATIONS
       The sound card should be	capable	of 44100 Hz  sampling  via  the	 sound
       drivers.	  You  must  use an external mixer program to select the input
       source device, level, etc.  Since these unknowns	affect the  amplitude,
       there is	no reference to	voltage	on the Y axis; it is in	fact, unknown.
       Instead	you're	given  the scale in pixels per sample unit.  Note that
       the serial oscilloscope devices don't have this limitation.  They  have
       real voltage labels on the Y axis.

       Signal  math  is	 only  valid if	Channel	1 and 2	contain	signals	of the
       same sampling rate.  It is up to	you to make sure this is the case. Do-
       ing math	on signals of different	sample rates  will  produce  incorrect
       results!

       The automatic measurements count	zero crossings and divide to determine
       the  frequency and period.  If these zero crossings are not "regularly-
       periodic", these	measurements could be invalid.	 Xoscope  does	under-
       stand  how  to  measure the built-in FFT	functions by locating the peak
       frequency.  Use manual cursor positioning to get	more precise  measure-
       ments.

       Your  sound card	is most-likely AC coupled so you will never see	any DC
       offset.	You probably can't get DC coupling by just shorting the	 input
       capacitors on your sound	card.  Use serial hardware to see DC offsets.

       The display may not be able to keep up if you give it too much to plot,
       depending  on your sound	card, graphics card, and processor speed.  Ex-
       ternal math commands are	particularly expensive since the  kernel  must
       then split the available	CPU cycles across multiple processes.  To max-
       imize refresh speed, hide all unneeded channels,	use point or point ac-
       cumulate	mode, zoom in on Sec/Div as much as possible, and turn off the
       graticule.

BUGS
       The keyboard interface may be confusing.

AUTHOR
       Oscope  was  written  by	 Tim  Witham (twitham@quiknet.com), originally
       based on	"scope"	by Jeff	Tranter	(Jeff_Tranter@Mitel.COM).  Most	recent
       work is by Brent	Baccala	(cosine@freesoft.org).	 Xoscope  is  released
       under  the conditions of	the GNU	General	Public License.	 See the files
       README and COPYING in the distribution for details.

Linux				 May  6	 2001			    XOSCOPE(1)